There is a lack of data on instrument dexterity and interface resistance with respect to the emerging surgical technology of LESS ports. A comparative analysis was conducted to characterize the force to maneuver laparoscopic instruments at various working angles within three commercially available LESS ports. A novel test fixture was created where working angles of the instruments were systematically varied in both the horizontal and sagittal plane within synthetic skin and rigid inserts. Two standard 5-mm laparoscopic graspers and a 10-mm simulated laparoscope were inserted into the trocars of the SILS™, TriPort™ and GelPOINT™ LESS ports. The positions of the laparoscope and grasper (G1) were fixed, while the working instrument’s position (G2) was systematically varied to create a range-of-motion. The static force required to maintain a specific position for G2 was measured using a digital force gauge for that range-of-motion. The resistance created by each LESS port was most noticeable at greater separation angles. The GelPOINT™ provided the least resistance to instrument movement; while the TriPort™ required the greatest amount of force at all angular positions. The 15-mm skin interface yielded lower overall resistance for all ports compared to the 30-mm skin interface. Resistance created by each LESS port increased with greater angular separation. Increased thickness and rigidity of the abdominal wall resulted in greater static forces and reduced instrument range-of-motion for all surgical ports. LESS port design and geometry heavily influenced overall instrument range-of-motion, as well as the resistance found at extreme separation angles. Surgeons should consider the degree of instrument motion required specific to the procedure being performed when selecting a LESS port.
Skip Nav Destination
hallbeck@unl.edu.
Article navigation
June 2012
Research Papers
Evaluation of Instrument Dexterity and Static Resistance of Laparoendoscopic Single-Site (LESS) Surgical Ports
Adam E. de Laveaga,
Adam E. de Laveaga
Department of Mechanical and Materials Engineering, University of Nebraska- Lincoln
, Lincoln, NE, 68588
Search for other works by this author on:
Bernadette McCrory,
Bernadette McCrory
Department of Mechanical and Materials Engineering, University of Nebraska- Lincoln
, Lincoln, NE, 68588
Search for other works by this author on:
Chad A. LaGrange,
Chad A. LaGrange
Department of Surgery, Section of Urology, University of Nebraska Medical Center
, Omaha, NE, 68198
Search for other works by this author on:
M. Susan Hallbeck
hallbeck@unl.edu.
M. Susan Hallbeck
Department of Mechanical and Materials Engineering, University of Nebraska- Lincoln
, Lincoln, NE, 68588
Search for other works by this author on:
Adam E. de Laveaga
Department of Mechanical and Materials Engineering, University of Nebraska- Lincoln
, Lincoln, NE, 68588
Bernadette McCrory
Department of Mechanical and Materials Engineering, University of Nebraska- Lincoln
, Lincoln, NE, 68588
Chad A. LaGrange
Department of Surgery, Section of Urology, University of Nebraska Medical Center
, Omaha, NE, 68198
M. Susan Hallbeck
Department of Mechanical and Materials Engineering, University of Nebraska- Lincoln
, Lincoln, NE, 68588hallbeck@unl.edu.
J. Med. Devices. Jun 2012, 6(2): 021002 (7 pages)
Published Online: April 6, 2012
Article history
Received:
June 12, 2011
Revised:
January 11, 2012
Published:
April 5, 2012
Online:
April 6, 2012
Citation
de Laveaga, A. E., McCrory, B., LaGrange, C. A., and Hallbeck, M. S. (April 6, 2012). "Evaluation of Instrument Dexterity and Static Resistance of Laparoendoscopic Single-Site (LESS) Surgical Ports." ASME. J. Med. Devices. June 2012; 6(2): 021002. https://doi.org/10.1115/1.4006130
Download citation file:
Get Email Alerts
Cited By
Quantifying Viscous Damping and Stiffness in Parkinsonism Using Data-Driven Model Estimation and Admittance Control
J. Med. Devices (December 2022)
Comparison of Temperature Equilibrium Rate and Cell Growth/Viability Under Air Circulation in Cryogenic Storage Container
J. Med. Devices (December 2022)
Preclinical Evaluation of a Task-Shifting Contraceptive Implant Insertion Device for Use in Low- and Middle-Income Countries
J. Med. Devices (December 2022)
Related Articles
Vaginal Approach for Uterus Separation During Laparoscopic Hysterectomy
J. Med. Devices (June,2010)
Design of an Instrument Guide for MRI-Guided Percutaneous Interventions
J. Med. Devices (June,2011)
Wrinkling of Plane Isotropic Biological Membranes
J. Appl. Mech (May,2007)
Design of a Tritium-In-Air Monitor Using Field-Programmable Gate Arrays
ASME J of Nuclear Rad Sci (October,2016)
Related Chapters
Modeling and Realization on a Chip of Growth Hormone Secretion Mechanism
Intelligent Engineering Systems Through Artificial Neural Networks, Volume 17
Health and Safety and Emergency Response
Pipeline Transportation of Carbon Dioxide Containing Impurities
Sunscreen or Smokescreen?
Hot Air Rises and Heat Sinks: Everything You Know about Cooling Electronics Is Wrong