Elevated stresses in deep plantar tissue of diabetic neuropathic patients were associated with an increased risk for foot ulceration, but only interfacial foot pressures are currently measured to evaluate susceptibility to ulcers. The goals of this study were to develop a real-time patient-specific plantar tissue stress monitor based on the Hertz contact theory. The biomechanical model for stress calculations considers the heel and metatarsal head pads, where most ulcers occur. For calculating stress concentrations around the bone-pad interface, plantar tissue is idealized as elastic and incompressible semi-infinite bulk (with properties measured by indentation), which is penetrated by a rigid sphere with the bone’s radius of curvature (from X-ray). Hertz’s theory is used to solve the bone-pad mechanical interactions, after introducing correction coefficients to consider large deformations. Foot-shoe forces are measured to solve and display the principal compressive, tensile, and von Mises plantar tissue stresses in real time. Our system can be miniaturized in a handheld computer, allowing plantar stress monitoring in the patient’s natural environment. Small groups of healthy subjects and diabetic patients participated in an evaluation study in which the differences between free walking and treadmill walking were examined. We also compared gait on a flat surface to gait on an ascending/descending slope of and when ascending/descending stairs. Peak internal compression stress was about threefold greater than the interface pressure at the calcaneus region. Subjects who were inexperience in treadmill walking displayed high gait-cycle variability in the internal stresses as well as poor foot loading. There was no statistical difference between gait on a flat surface and gait when ascending/descending a slope. Internal stresses under the calcaneus during gait on a flat surface, however, were significantly higher than when ascending/descending stairs. We conclude that the present stress monitor is a promising tool for real-time patient-specific evaluation of deep tissue stresses, providing valuable information in the effort to protect diabetic patients from foot ulceration. Clinical studies are now underway to identify which stress parameters can distinguish between diabetic and normal subjects; these parameters may be used for establishing injury threshold criteria.
Skip Nav Destination
gefen@eng.tau.ac.ul
Article navigation
March 2008
Research Papers
The Diabetic Foot Load Monitor: A Portable Device for Real-Time Subject-Specific Measurements of Deep Plantar Tissue Stresses During Gait
Eran Atlas,
Eran Atlas
Department of Biomedical Engineering,
Tel Aviv University
, Tel Aviv 69978, Israel
Search for other works by this author on:
Ziva Yizhar,
Ziva Yizhar
Department of Physical Therapy,
Tel Aviv University
, Tel Aviv 69978, Israel
Search for other works by this author on:
Amit Gefen
Amit Gefen
Department of Biomedical Engineering,
gefen@eng.tau.ac.ul
Tel Aviv University
, Tel Aviv 69978, Israel
Search for other works by this author on:
Eran Atlas
Department of Biomedical Engineering,
Tel Aviv University
, Tel Aviv 69978, Israel
Ziva Yizhar
Department of Physical Therapy,
Tel Aviv University
, Tel Aviv 69978, Israel
Amit Gefen
Department of Biomedical Engineering,
Tel Aviv University
, Tel Aviv 69978, Israelgefen@eng.tau.ac.ul
J. Med. Devices. Mar 2008, 2(1): 011005 (10 pages)
Published Online: March 10, 2008
Article history
Received:
May 31, 2007
Revised:
February 11, 2008
Published:
March 10, 2008
Citation
Atlas, E., Yizhar, Z., and Gefen, A. (March 10, 2008). "The Diabetic Foot Load Monitor: A Portable Device for Real-Time Subject-Specific Measurements of Deep Plantar Tissue Stresses During Gait." ASME. J. Med. Devices. March 2008; 2(1): 011005. https://doi.org/10.1115/1.2891241
Download citation file:
Get Email Alerts
Cited By
Elliptical Trainer Redesign to Diversify Muscles Recruitment
J. Med. Devices (September 2022)
High-Speed Jet Injector for Pharmaceutical Applications
J. Med. Devices (September 2022)
DynaRing: A Patient-Specific Mitral Annuloplasty Ring With Selective Stiffness Segments
J. Med. Devices (September 2022)
Scale Effects on Performance of BLDC Micromotors for Internal Biomedical Applications: A Finite Element Analysis
J. Med. Devices (September 2022)
Related Articles
A Minimally Invasive Device for Continuous Glucose Monitoring in Infants
J. Med. Devices (June,2008)
Cerebrospinal Fluid Volume Monitoring for Hydrocephalus Therapy
J. Med. Devices (June,2011)
Design of a Pressure Measuring Syringe
J. Med. Devices (June,2010)
Vibrotactile Balance Rehabilitation Gait Assist Device
J. Med. Devices (June,2009)
Related Proceedings Papers
Related Chapters
Theoretical Analysis and Application of Pile Shaft Resistance Strengthening Effect
International Conference on Optimization Design (ICOD 2010)
Introduction and scope
Impedimetric Biosensors for Medical Applications: Current Progress and Challenges
Conclusion & executive summary
Photodynamic Therapy Mediated by Fullerenes and their Derivatives