Readily available store brand, or “home,” thermometers are used countless times in the home and clinic as a first diagnostic measure of body temperature. Measurement inaccuracies may lead to unnecessary medical visits or medication (false positives), or, potentially worse, lack of intervention when a person is truly sick (false negatives). A critical first step in the design process is to determine the shortcomings of the existing designs. For this project, we evaluated the accuracy of three currently available store brand thermometers in a pediatric population. The accuracies of the thermometers were assessed by comparing their body temperature predictions to those measured by a specially designed and calibrated and fast-responding reference thermometer. The reference thermometer was placed at the measurement site simultaneously with the store brand thermometer and recorded the temperature at the measurement site continuously. More than 300 healthy or sick pediatric subjects were enrolled in this study. Temperatures were measured at both the oral and axillary (under the arm) sites. The store brand thermometer measurements characteristically deviated from the reference thermometer temperature after 120 s, and the deviations did not follow a consistent pattern. The Brand C thermometers had the greatest deviations of up to 3.7 °F (2.1 °C), while the Brand A thermometers had the lowest deviations; however, they still deviated by up to 1.9 °F (1.1 °C). The data showed that the tested store brand thermometers had lower accuracy than the ±0.2 °F (0.1 °C) indicated in their Instructions for Use. Our recorded reference (transient) data showed that there was a wide variation in the transient temperature profiles. The store brand thermometers tested stated in their documentation that they are able to predict a body temperature based on transient temperature values over the first 5–10 s of measurements, implying that they use an embedded algorithm to extrapolate to the steady-state temperature. Significant deviations from the maximum temperature after time t = 4.6t0.63 illustrated that the transient temperature profiles may not be represented by an exponential function with a single time constant, t0.63. The accuracy of those embedded algorithms was not confirmed by our study, since the predicted body temperatures do not capture the large variations observed over the initial 10 s of the measurements. A thermometer with an error of several degrees Fahrenheit may result in a false positive or negative diagnosis of fever in children. The transient temperature measurements from our clinical study represent unique and critical data for helping to design the next generation of readily available, highly accurate, home thermometers.
Skip Nav Destination
Article navigation
March 2019
Research-Article
Identifying Critical Design Parameters for Improved Body Temperature Measurements: A Clinical Study Comparing Transient and Predicted Temperature Measurements
Oleg Vesnovsky,
Oleg Vesnovsky
FDA Center for Devices & Radiological Health,
Silver Spring, MD 20993
e-mail: oleg.vesnovsky@fda.hhs.gov
Silver Spring, MD 20993
e-mail: oleg.vesnovsky@fda.hhs.gov
Search for other works by this author on:
Liang Zhu,
Liang Zhu
Department of Mechanical Engineering,
University of Maryland,
Baltimore County,
Baltimore, MD 21250
University of Maryland,
Baltimore County,
Baltimore, MD 21250
Search for other works by this author on:
Laurence W. Grossman,
Laurence W. Grossman
FDA Center for Devices & Radiological Health,
Silver Spring, MD 20993
Silver Spring, MD 20993
Search for other works by this author on:
Jon P. Casamento,
Jon P. Casamento
FDA Center for Devices & Radiological Health,
Silver Spring, MD 20993
Silver Spring, MD 20993
Search for other works by this author on:
Alireza Chamani,
Alireza Chamani
Department of Mechanical Engineering,
University of Maryland,
Baltimore County,
Baltimore, MD 21250
University of Maryland,
Baltimore County,
Baltimore, MD 21250
Search for other works by this author on:
Nadeesri Wijekoon,
Nadeesri Wijekoon
Department of Mathematics and Statistics,
University of Maryland,
Baltimore County,
Baltimore, MD 21250
University of Maryland,
Baltimore County,
Baltimore, MD 21250
Search for other works by this author on:
L. D. Timmie Topoleski
L. D. Timmie Topoleski
FDA Center for Devices & Radiological Health,
Silver Spring, MD 20993;
Department of Mechanical Engineering,
University of Maryland,
Baltimore County,
Baltimore, MD 21250
e-mail: topoleski@umbc.edu
Silver Spring, MD 20993;
Department of Mechanical Engineering,
University of Maryland,
Baltimore County,
Baltimore, MD 21250
e-mail: topoleski@umbc.edu
Search for other works by this author on:
Oleg Vesnovsky
FDA Center for Devices & Radiological Health,
Silver Spring, MD 20993
e-mail: oleg.vesnovsky@fda.hhs.gov
Silver Spring, MD 20993
e-mail: oleg.vesnovsky@fda.hhs.gov
Liang Zhu
Department of Mechanical Engineering,
University of Maryland,
Baltimore County,
Baltimore, MD 21250
University of Maryland,
Baltimore County,
Baltimore, MD 21250
Laurence W. Grossman
FDA Center for Devices & Radiological Health,
Silver Spring, MD 20993
Silver Spring, MD 20993
Jon P. Casamento
FDA Center for Devices & Radiological Health,
Silver Spring, MD 20993
Silver Spring, MD 20993
Alireza Chamani
Department of Mechanical Engineering,
University of Maryland,
Baltimore County,
Baltimore, MD 21250
University of Maryland,
Baltimore County,
Baltimore, MD 21250
Nadeesri Wijekoon
Department of Mathematics and Statistics,
University of Maryland,
Baltimore County,
Baltimore, MD 21250
University of Maryland,
Baltimore County,
Baltimore, MD 21250
L. D. Timmie Topoleski
FDA Center for Devices & Radiological Health,
Silver Spring, MD 20993;
Department of Mechanical Engineering,
University of Maryland,
Baltimore County,
Baltimore, MD 21250
e-mail: topoleski@umbc.edu
Silver Spring, MD 20993;
Department of Mechanical Engineering,
University of Maryland,
Baltimore County,
Baltimore, MD 21250
e-mail: topoleski@umbc.edu
1O. Vesnovsky and L. Zhu both contributed equally.
2Corresponding authors.
Manuscript received May 16, 2017; final manuscript received September 24, 2018; published online December 4, 2018. Assoc. Editor: Xiaoming He. This work is in part a work of the U.S. Government. ASME disclaims all interest in the U.S. Government's contributions.
J. Med. Devices. Mar 2019, 13(1): 011005 (15 pages)
Published Online: December 4, 2018
Article history
Received:
May 16, 2017
Revised:
September 24, 2018
Citation
Vesnovsky, O., Zhu, L., Grossman, L. W., Casamento, J. P., Chamani, A., Wijekoon, N., and Timmie Topoleski, L. D. (December 4, 2018). "Identifying Critical Design Parameters for Improved Body Temperature Measurements: A Clinical Study Comparing Transient and Predicted Temperature Measurements." ASME. J. Med. Devices. March 2019; 13(1): 011005. https://doi.org/10.1115/1.4041589
Download citation file:
159
Views
0
Citations
Get Email Alerts
Cited By
Stakeholders, Prototypes, and Settings of Front-End Medical Device Design Activities
J. Med. Devices (September 2022)
DynaRing: A Patient-Specific Mitral Annuloplasty Ring With Selective Stiffness Segments
J. Med. Devices (September 2022)
Scale Effects on Performance of BLDC Micromotors for Internal Biomedical Applications: A Finite Element Analysis
J. Med. Devices (September 2022)
Elliptical Trainer Redesign to Diversify Muscles Recruitment
J. Med. Devices
Related Articles
The Use of a High Temperature Wind Tunnel for MT-SOFC Testing—Part I: Detailed Experimental Temperature Measurement of an MT-SOFC Using an Avant-Garde High Temperature Wind Tunnel and Various Measurement Techniques
J. Fuel Cell Sci. Technol (December,2010)
Design and Test of Carbon Nanotube Biwick Structure for High-Heat-Flux Phase Change Heat Transfer
J. Heat Transfer (May,2010)
Lagging Behavior in Biological Systems
J. Heat Transfer (May,2012)
Thin Film Conducting and Semiconducting Resistance Thermometers for Surface Temperature Measurement
J. Eng. Power (July,1977)
Related Proceedings Papers
Related Chapters
EVALUATION OF THERMAL PROBE COOLING EFFECT BASED ON PIPELINE PARALLEL OPTICAL CABLE
Pipeline Integrity Management Under Geohazard Conditions (PIMG)
Transient Temperature and Heat Flux Measurement Using Thin-Film Microsensors
Ultrasonic Welding of Lithium-Ion Batteries
Even a Watched Pot Boils Eventually
Hot Air Rises and Heat Sinks: Everything You Know about Cooling Electronics Is Wrong