Skip to Main Content
Skip Nav Destination
ASME Press Select Proceedings
Proceedings of the 10th International Symposium on Cavitation (CAV2018)
Editor
Joseph Katz
Joseph Katz
Search for other works by this author on:
ISBN:
9780791861851
No. of Pages:
1108
Publisher:
ASME Press
Publication date:
2018

Cavitation has been observed in the cone of a syringe actuated by an autoinjector device. Numerical simulations were used to determine if the cone can enhance the collapse of a bubble. We found the collapse of a bubble in a cone can be more violent than the collapse of a bubble close to a flat wall or in a free space due to the reflected wave focusing on the axis of symmetry.

Introduction
Numerical Setup
Results & Discussion
Conclusion
References
This content is only available via PDF.
Close Modal

or Create an Account

Close Modal
Close Modal