The main biomechanical factors which may affect the accuracy of the oscillometric method for indirect blood pressure measurement are analyzed using a new model of brachial hemodynamics. In a first stage of this work, the model has been used to reproduce some well-known responses of collapsing arteries, such as the sharp increase in compliance, and the nonlinear pressure-flow characteristic with negative dynamic resistance. In a second stage the model has been linked to the arm tissue mechanics description presented in a previous work. The final model so obtained has then been employed to analyze the pattern of the main hemodynamic quantities (pressure pulsations in the cuffs, blood volume changes, blood flow upstream and downstream of the cuffs) during deflation manoeuvres. The simulation results agree with those found in the recent literature quite well. Results indicate that the cuff pressure value for maximum pulsations exhibits a large plateau, located approximately around the mean arterial pressure. However, stiffness of wall artery, or stretching of the cuff internal surface, may significantly alter the obtained results causing a phenomena of “pseudohypertension.”
Article navigation
Technical Papers
Mathematical Modeling of Noninvasive Blood Pressure Estimation Techniques—Part II: Brachial Hemodynamics
Mauro Ursino
,
Mauro Ursino
Department of Electronics, Computer Science and Systems, University of Bologna, viale Risorgimento 2, l 40136, Bologna, Italy
Search for other works by this author on:
Cristina Cristalli
Cristina Cristalli
Department of Electronics and Automatics, University of Ancona, via Brecce Bianche, 160131 Ancona, Italy
Search for other works by this author on:
Mauro Ursino
Department of Electronics, Computer Science and Systems, University of Bologna, viale Risorgimento 2, l 40136, Bologna, Italy
Cristina Cristalli
Department of Electronics and Automatics, University of Ancona, via Brecce Bianche, 160131 Ancona, Italy
J Biomech Eng. Feb 1995, 117(1): 117-126 (10 pages)
Published Online: February 1, 1995
Article history
Received:
April 9, 1993
Revised:
April 20, 1994
Online:
October 30, 2007
Citation
Ursino, M., and Cristalli, C. (February 1, 1995). "Mathematical Modeling of Noninvasive Blood Pressure Estimation Techniques—Part II: Brachial Hemodynamics." ASME. J Biomech Eng. February 1995; 117(1): 117–126. https://doi.org/10.1115/1.2792259
Download citation file:
- Ris (Zotero)
- Reference Manager
- EasyBib
- Bookends
- Mendeley
- Papers
- EndNote
- RefWorks
- BibTex
- ProCite
- Medlars
Close
Sign In
Get Email Alerts
Cited By
Reliability-Based Design Optimization of a Cemented Prosthesis in a Femur Undergoing Bone Remodeling
J Biomech Eng (April 2020)
Automated Quantification of Diseased Thoracic Aortic Longitudinal Centerline and Surface Curvatures
J Biomech Eng (April 2020)
ANNUAL SPECIAL ISSUE “Biomechanical Engineering—2019 Year in Review”
J Biomech Eng (March 2020)
Related Articles
Possible Sources of Discrepancy Between Sphygmomanometer Cuff Pressure and Blood Pressure Quantified in a Collapsible-Tube Analogue
J Biomech Eng (February, 1992)
A Computational Method for Analyzing the Biomechanics of Arterial Bruits
J Biomech Eng (May, 2017)
Modeling the Effect of Red Blood Cells Deformability on Blood Flow Conditions in Human Carotid Artery Bifurcation
J Biomech Eng (January, 2017)
Related Proceedings Papers
Related Chapters
Introduction and Scope
High Frequency Piezo-Composite Micromachined Ultrasound Transducer Array Technology for Biomedical Imaging
Tissue and blood-material interactions
Biocompatible Nanomaterials for Targeted and Controlled Delivery of Biomacromolecules
Bioinformatic Analysis of the Malay Peninsular Macaca Fascicularis Based on Cytochrome C Oxidase I Sequences
International Conference on Computer Research and Development, 5th (ICCRD 2013)