We have constructed an electrokinetic surface probe capable of applying small sinusoidal currents to the surface of articular cartilage and measuring the resulting current-generated stress with a piezoelectric sensor. Using the probe, we have characterized the electromechanical response of excised discs of normal and chemically modified adult bovine femoropatellar groove cartilage. The measured stress amplitude was proportional to the applied current density and inversely proportional to the excitation frequency, consistent with a poroelastic model. As a function of bath pH, the stress amplitude exhibited a minimum in the range pH 2.4–2.8 and the phase underwent an abrupt 180° transition in the same range, consistent with an electrokinetic mechanism as the origin of the current-generated mechanical stress. Digestion of the tissue with trypsin resulted in a progressive loss of highly charged proteoglycan molecules from the tissue, with a concomitant decrease in the measured stress amplitude. These results support the feasibility of surface measurements as a means of assessing electromechanical transduction in cartilage and of detecting subtle molecular-level degradative changes in the extracellular matrix. This technique of surface spectroscopy provides a new means of nondestructively measuring the material properties of cartilage on intact joints and detecting degradative changes such as those seen in the earliest stages of osteoarthritis.
Skip Nav Destination
Article navigation
November 1994
Research Papers
Nondestructive Detection of Cartilage Degeneration Using Electromechanical Surface Spectroscopy
Scott I. Berkenblit,
Scott I. Berkenblit
Continuum Electromechanics Group, Laboratory for Electromagnetic and Electronic Systems, Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Cambridge, MA 02139; Harvard-M.I.T. Division of Health Sciences and Technology
Search for other works by this author on:
Eliot H. Frank,
Eliot H. Frank
Continuum Electromechanics Group, Laboratory for Electromagnetic and Electronic Systems, Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Cambridge, MA 02139; Harvard-M.I.T. Division of Health Sciences and Technology
Search for other works by this author on:
Evan P. Salant,
Evan P. Salant
Continuum Electromechanics Group, Laboratory for Electromagnetic and Electronic Systems, Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Cambridge, MA 02139; Harvard-M.I.T. Division of Health Sciences and Technology
Search for other works by this author on:
Alan J. Grodzinsky
Alan J. Grodzinsky
Continuum Electromechanics Group, Laboratory for Electromagnetic and Electronic Systems, Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Cambridge, MA 02139; Harvard-M.I.T. Division of Health Sciences and Technology
Search for other works by this author on:
Scott I. Berkenblit
Continuum Electromechanics Group, Laboratory for Electromagnetic and Electronic Systems, Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Cambridge, MA 02139; Harvard-M.I.T. Division of Health Sciences and Technology
Eliot H. Frank
Continuum Electromechanics Group, Laboratory for Electromagnetic and Electronic Systems, Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Cambridge, MA 02139; Harvard-M.I.T. Division of Health Sciences and Technology
Evan P. Salant
Continuum Electromechanics Group, Laboratory for Electromagnetic and Electronic Systems, Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Cambridge, MA 02139; Harvard-M.I.T. Division of Health Sciences and Technology
Alan J. Grodzinsky
Continuum Electromechanics Group, Laboratory for Electromagnetic and Electronic Systems, Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Cambridge, MA 02139; Harvard-M.I.T. Division of Health Sciences and Technology
J Biomech Eng. Nov 1994, 116(4): 384-392 (9 pages)
Published Online: November 1, 1994
Article history
Received:
March 23, 1993
Revised:
October 28, 1993
Online:
March 17, 2008
Citation
Berkenblit, S. I., Frank, E. H., Salant, E. P., and Grodzinsky, A. J. (November 1, 1994). "Nondestructive Detection of Cartilage Degeneration Using Electromechanical Surface Spectroscopy." ASME. J Biomech Eng. November 1994; 116(4): 384–392. https://doi.org/10.1115/1.2895788
Download citation file:
Get Email Alerts
Effect of Internal Mechanical Environment of Porous Scaffolds on Mechano-driven Bone Ingrowth: A Numerical Study
J Biomech Eng (September 2023)
In Silico Mechanical Effort Analysis of the All-On-4 Design Performed With Platform-Switching Distal Short Dental Implants
J Biomech Eng (September 2023)
Related Articles
Contact Creep of Biphasic Cartilage Layers
J. Appl. Mech (March,1999)
Transient and Cyclic Responses of Strain-Generated Potential in Rabbit Patellar Tendon Are Frequency and pH Dependent
J Biomech Eng (October,2000)
A Comparison Between Mechano-Electrochemical and Biphasic Swelling Theories for Soft Hydrated Tissues
J Biomech Eng (February,2005)
Related Proceedings Papers
Related Chapters
Effects of Ultrasound Stimulation on Chondrocytes in Three-Dimensional Culture in Relation to the Production of Regenerative Cartilage Tissue
Biomedical Applications of Vibration and Acoustics in Therapy, Bioeffect and Modeling
Essential Probability and Statistics
Nonlinear Regression Modeling for Engineering Applications: Modeling, Model Validation, and Enabling Design of Experiments
Principles
Ultrasonic Methods for Measurement of Small Motion and Deformation of Biological Tissues for Assessment of Viscoelasticity