An investigation was performed to determine the effects of the presence of two lengths of proximal Mu¨ller prosthesis on predicted failure loads, as compared to those for an intact femur. Three-dimensional stresses in a bone/cement/prosthesis system were determined using finite element methods, with both isotropic and transversely isotropic material properties used for the diaphyseal cortex. Significant increases in prosthesis stem stresses were found when the transversely isotropic material properties were employed in the diaphyseal cortex. This leads to the conclusion that accurate anisotropic material properties for bone are essential for precise stress determination and optimum design in prosthetic implants. Failure loads were also predicted for vertical compression and axial torque, similar to available experimental conditions, and were within the range of the experimental failure data found in the literature. The technique developed herein can be used to systematically assess existing as well as future implant designs, taking into account the complex three-dimensional interaction effects of the overall bone/cement/prosthesis system.
Skip Nav Destination
Article navigation
February 1986
Research Papers
Stress Analysis and Failure Prediction in the Proximal Femur Before and After Total Hip Replacement
H. H. Vichnin,
H. H. Vichnin
General Electric Space Division, Valley Forge, Pa.
Search for other works by this author on:
S. C. Batterman
S. C. Batterman
Department of Bioengineering, University of Pennsylvania, Philadelphia, Pa. 19104
Search for other works by this author on:
H. H. Vichnin
General Electric Space Division, Valley Forge, Pa.
S. C. Batterman
Department of Bioengineering, University of Pennsylvania, Philadelphia, Pa. 19104
J Biomech Eng. Feb 1986, 108(1): 33-41 (9 pages)
Published Online: February 1, 1986
Article history
Received:
May 22, 1983
Revised:
August 27, 1985
Online:
June 12, 2009
Citation
Vichnin, H. H., and Batterman, S. C. (February 1, 1986). "Stress Analysis and Failure Prediction in the Proximal Femur Before and After Total Hip Replacement." ASME. J Biomech Eng. February 1986; 108(1): 33–41. https://doi.org/10.1115/1.3138577
Download citation file:
Get Email Alerts
Simulating the Growth of TATA-Box Binding Protein-Associated Factor 15 Inclusions in Neuron Soma
J Biomech Eng (December 2024)
Effect of Structure and Wearing Modes on the Protective Performance of Industrial Safety Helmet
J Biomech Eng (December 2024)
Sex-Based Differences and Asymmetry in Hip Kinematics During Unilateral Extension From Deep Hip Flexion
J Biomech Eng (December 2024)
Related Articles
Survey of Stress Analyses of the Femoral Hip Prosthesis
Appl. Mech. Rev (January,2000)
Effect of Geometrical Uncertainty on Cemented Hip Implant Structural Integrity
J Biomech Eng (May,2009)
Physically Realizable Three-Dimensional Bone Prosthesis Design With Interpolated Microstructures
J Biomech Eng (March,2017)
Design of an Endoreactor for the Cultivation of a Joint-Like-Structure
J. Med. Devices (June,2009)
Related Proceedings Papers
Related Chapters
Introduction and Definitions
Handbook on Stiffness & Damping in Mechanical Design
Subsection NB—Class 1 Components
Companion Guide to the ASME Boiler & Pressure Vessel Code, Volume 1, Second Edition
Development and Structure of the German Common Cause Failure Data Pool (PSAM-0020)
Proceedings of the Eighth International Conference on Probabilistic Safety Assessment & Management (PSAM)