In order to determine the effect of application pressure on the accuracy of skin temperature measurements for area contact sensors, low values of pressure (2-20 mm Hg) were applied to the mid-thigh and to the lateral aspect of the trochanter of human subjects using a thin, circular disk with a thermistor mounted in the base. From measurements of the local skin temperatures, it was determined that a pressure of 2 mm Hg is adequate to measure the skin temperature accurately. Applying larger pressure results in higher local skin temperatures with the thighs showing larger temperature increases than the trochanters. The results of a finite difference analysis indicate that the increases in skin temperature at higher pressures can be accounted for by the physical phenomena associated with the penetration of the sensor into the tissue. After the release of pressure, the local skin temperature immediately decreased for all subjects indicating little or no reactive hyperemia was occurring. A method of compensating for the changes in local skin temperature which are due to whole body transient thermal effects was also developed. Use of this method allows the effects of the local pressure application to be separated from the transient environmental effects.
Skip Nav Destination
Article navigation
November 1979
Research Papers
The Effect of Pressure on Skin Temperature Measurements for a Disk Sensor
S. D. Mahanty,
S. D. Mahanty
California State University, Fresno, Calif.
Search for other works by this author on:
R. B. Roemer
R. B. Roemer
Department of Mechanical and Environmental Engineering, University of California, Santa Barbara, Calif. 93106; Memorial Rehabilitation Foundation Center, Santa Barbara, Calif. 93105
Search for other works by this author on:
S. D. Mahanty
California State University, Fresno, Calif.
R. B. Roemer
Department of Mechanical and Environmental Engineering, University of California, Santa Barbara, Calif. 93106; Memorial Rehabilitation Foundation Center, Santa Barbara, Calif. 93105
J Biomech Eng. Nov 1979, 101(4): 261-266 (6 pages)
Published Online: November 1, 1979
Article history
Received:
October 30, 1978
Revised:
July 16, 1979
Online:
October 21, 2010
Citation
Mahanty, S. D., and Roemer, R. B. (November 1, 1979). "The Effect of Pressure on Skin Temperature Measurements for a Disk Sensor." ASME. J Biomech Eng. November 1979; 101(4): 261–266. https://doi.org/10.1115/1.3426255
Download citation file:
Get Email Alerts
Cited By
Related Articles
Cryosurgery of Normal and Tumor Tissue in the Dorsal Skin Flap Chamber: Part I—Thermal Response
J Biomech Eng (August,2001)
Numerical and Experimental Simulations as Symbiotic Tools for Solving Complex Biothermal Problems
J. Med. Devices (June,2010)
Perspective in Nanoneural Electronic Implants With Wireless Power-Feed and Sensory Control
J. Nanotechnol. Eng. Med (May,2010)
Modeling the Thermal Responses of the Skin Surface During Hand-Object Interactions
J Biomech Eng (April,2008)
Related Proceedings Papers
Related Chapters
Modeling and Realization on a Chip of Growth Hormone Secretion Mechanism
Intelligent Engineering Systems Through Artificial Neural Networks, Volume 17
Example Open-Book Questions
Quick Guide to the API 570 - Certified Pipework Inspector Syllabus
Sunscreen or Smokescreen?
Hot Air Rises and Heat Sinks: Everything You Know about Cooling Electronics Is Wrong