A preliminary, prospective, multicenter study using the 3.5 mm low contact dynamic compression plate (LCDCP) for traumatic and reconstructive problems in the upper extremity skeleton is presented. Over a 13-month period in three institutions, 26 plates were implanted in 21 patients. Twelve were male and 9 female, with an average age of 26 years. Eleven LCDCP were used for acute fractures. Fifteen plates were used in reconstructions, including nonunions; radial osteotomies; and wrist arthrodeses. Twelve were stainless steel and 14 titanium. Plate lengths ranged from 6 holes to 10 holes. Multi-directional contouring was required in 6 cases. Obliquely placed 3.5 mm design “shaft” lag screws were used in 2 cases. In 4 cases, the plate design permitted compression in 2 opposing directions along the longitudinal axis of the bone. In one case, the newer hole design allowed placement of an oblique lag screw at a 40-degree angle to the plate. All patients were treated with a functional aftercare program. Radiographic and clinical union was achieved within 3 months in each case. There were no immediate or late postoperative problems. Functional limb recovery was noted in each case. One patient had plate removal 6 months post wrist arthrodesis due to extensor tendon inflammation. The plate was securely fixed. Microscopic sections of zones of contact and non-contact clearly demonstrated increased osteoporosis in the zones of plate contact. This experience supports the fact that with the LCDCP internal fixation is secure, plate design allows easier 3-dimensional contouring, and screw placement is facilitated.

1.
Perren
,
S.M.
,
Cordey
,
J.
,
Rahn
,
B.A.
,
Gauthier
,
E.
,
Schneider
,
E.
:
Early Temporary Porosis of Bone Induced by Internal Fixation Implants: A Reaction to Necrosis, not to Stress Protection
.
Clinical Orthopaedics and Related Res
.
232
: 139–151,
1988
.
2.
Swiontkowski
,
M.F.
,
Senft
,
D.
,
Taylor
,
S.
,
Agnew
S.G.
,
Santoro
,
V.
:
Plate Design has an Effect on Cortical Bone Perfusion
.
Proceedings of the Orthopaedic Research Society
.
66
,
1991
.
3.
Gasser
,
B.
,
Perren
,
S.M.
,
Schneider
,
E.
:
Parametric Numerical Design Optimization of Internal Fixation Plates
. Presented at 7th Meeting of the European Society of Biomechanics.
Aarhus, Denmark
,
1990
.
4.
Mears
,
D.C.
:
Materials and Orthopaedic Surgery
.
Baltimore
,
Williams & Wilkins
,
1979
.
5.
Lane
,
W.A.
:
The Operative Treatment of Fractures
. 2nd Edition,
London Medical Publishing
,
1914
.
6.
Lane
,
W.A.
:
The Direct Fixation of Fractures (abstract)
.
Transactions of Clinical Society London
27
: 167,
1894
.
7.
Platt
,
H.
:
Modern Trends in Orthopaedic Surgery
.
Medical Book Department of Harper and Brothers
. 29–33,
1950
.
8.
Charnley
,
J.C.
:
Positive Pressure in Arthrodesis of the Knee Joint
.
Journal of Bone and Joint Surgery
30-B
: 478–486,
1948
.
9.
Key
,
J.A.
:
Positive Pressure in Arthrodesis for Tuberculosis of the Knee Joint
.
Southern Med J
25
: 909–915,
1932
.
10.
Danis
,
R.
:
Theorie et practique de l'osteosythese
.
Paris
,
Masson
,
1949
.
11.
Eggers
,
G.W.
:
Internal Contact Splint
.
Journal of Bone and Joint Surgery
30A
, 40–52,
1948
.
12.
Bagby
,
G.W.
,
Janes
,
J.M.
:
The Effect of Compression Plates on the Rate of Fracture Healing Using a Special Plate
.
Am J Surg
95
: 761–771,
1958
.
13.
Allgower
,
M.
,
Perren
,
S.
,
Matter
,
P.
:
A New Plate Design: The Dynamic Compression Plate
.
Injury
2
: 40–47,
1970
.
14.
Uhthoff
,
H.K.
,
Dubuc
,
F.L.
:
Bone Structure Changes in the Dog Under Rigid Internal Fixation
.
Clinical Orthopaedics and Related Research
.
81
: 165–170,
1971
.
15.
Uhthoff
,
H.K.
,
Bardos
,
D.I.
,
Liskova-Kiar
M.
:
The Advantages of Titanium Alloy Over Stainless steel Plates for the Internal Fixation of Fractures: An Experimental Study in Dogs
.
Journal of Bone and Joint Surgery
63B
: 427–434,
1981
.
16.
DeLuca
,
P.A.
,
Lindsey
,
R.W.
,
Ruwe
,
P.A.
:
Refracture of Bones of the Forearm after the Removal of Compression Plates
.
Journal of Bone and Joint Surgery
70A
: 1372–1376,
1988
.
17.
Holzach
,
P.
,
Matter
,
P.
:
The Comparison of Steel and Titanium Dynamic Compression Plates Used for Internal Fixation of 256 Fractures of the Tibia
.
Injury
10
: 120–123,
1978
.
This content is only available via PDF.