Skip to Main Content
Skip Nav Destination
ASTM Selected Technical Papers
Biomaterials' Mechanical Properties
By
HE Kambic, M.S.
HE Kambic, M.S.
1
The Cleveland Clinic Foundation
?
Cleveland, OH symposium co-chair and co-editor
Search for other works by this author on:
AT Yokobori, Jr, Ph.D. Jr
AT Yokobori, Jr, Ph.D. Jr
2
Tohoku University
,
Sendai,
Japan
symposium co-chair and co-editor
Search for other works by this author on:
ISBN-10:
0-8031-1894-5
ISBN:
978-0-8031-1894-2
No. of Pages:
308
Publisher:
ASTM International
Publication date:
1994

A recent advancement in the design of the total artificial hip is the introduction of modularity. Within the last year clinical retrievals have shown that significant corrosion can occur at the cone-taper interface in some variations of the modular hip. The hypothesis of this study was that the stability of the cone-taper interface affects the amount of fretting corrosion that occurs. The development of a method that can study the taper corrosion and determine the most stable design is therefore necessary.

Design characteristics that may affect corrosion include the use of dissimilar metals, taper angle, taper diameter, percent coverage, machining tolerance, and head neck extension. The fretting that occurred at the interface was initially studied by measuring potential changes while cyclically loading hips in a universal testing machine. This method of measurement proved ineffective at isolating the fretting at the taper interface so current changes were monitored instead. The current was measured when the saline was below the crevice and when it was just above the crevice, and the two measurements were subtracted from each other to produce the taper fretting current.

When comparing two different designs, one showed a significant taper fretting current while the other's was minimal. In a second study, when comparing two different head neck extensions placed on identical stem designs, a significant taper fretting current was seen with the + 10-mm head as compared to a +0-mm head. It is concluded that this method can be used in the evaluation of designs to optimize stability and reduce the resulting fretting corrosion.

1.
Fontana
,
M. G.
,
Corrosion Engineering: Third Edition
,
McGraw-Hill Publishing Company
,
New York
,
1986
, pp. 51–9.
2.
Rostoker
,
W.
,
Galante
,
J. O.
, and
Lereim
,
P.
, “
Evaluation of Couple/Crevice Corrosion by Prosthetic Alloys Under In Vivo Conditions
,”
Journal of Biomedical Materials Research
, Vol.
12
,
1978
, pp. 823–29.
3.
Levine
,
D. L.
and
Staehl
,
R. W.
, “
Crevice Corrosion in Orthopedic Implant Metals
,
Journal of Biomedical Materials Research
, Vol.
11
,
1977
, pp. 553–61.
4.
Rose
,
R. M.
,
Schiller
,
A. L.
, and
Radin
,
E. L.
, “
Corrosion-Accelerated Mechanical Failure of a Vital-hum Nail-Plate
,”
Journal of Bone and Joint Surgery
, Vol.
54A
, No.
4
,
06
1972
, pp. 854–862.
5.
Kummer
,
F. J.
and
Rose
,
R. M.
, “
Corrosion of Titanium/Cobalt-Chromium Alloy Couples
,”
Journal of Bone and Joint Surgery
, Vol.
65A
, No.
8
,
10
1983
, pp. 1125–6.
6.
Rostoker
,
W.
,
Pretzel
,
C. W.
, and
Galante
,
J. O.
, “
Couple Corrosion Among Alloys for Skeletal Pros-theses
,”
Journal of Biomedical Materials Research
, Vol.
8
,
1974
, pp. 407–19.
7.
Lucas
,
L. C.
,
Buchanan
,
R. A.
, and
Lemons
,
J. E.
, “
Investigations on the Galvanic Corrosion of Multialloy Total Hip Prostheses
,”
Journal of Biomedical Materials Research
, Vol.
15
,
1981
, pp. 731–747.
8.
Sury
,
P.
and
Semlitsch
,
M.
, “
Corrosion Behavior of Cast and Forged Cobalt-Based Alloys for Double-Alloy Joint Endoprostheses
,”
Journal of Biomedical Materials Research
, Vol.
12
,
1978
, pp. 723–41.
9.
Brown
,
S. A.
and
Simpson
,
J. P.
, “
Crevice and Fretting Corrosion of Stainless-Steel Plates and Screws
,”
Journal of Biomedical Materials Research
, Vol.
15
,
1981
, pp. 867–78.
10.
Yao
,
L. A.
,
Gan
,
F. X.
,
Zhao
,
Y. X.
, et al
, “
Microelectrode Monitoring the Crevice Corrosion of Titanium
,”
Corrosion
, Vol.
47
, No.
6
, June
1991
, pp. 420–3.
11.
Mathiesen
,
E. B.
,
Lindgren
,
J. U.
,
Blomgren
,
G. G. A.
, and
Reinholt
,
F. P.
, “
Corrosion of Modular Hip Prostheses
,”
Journal of Bone and Joint Surgery
, Vol.
73B
, No.
4
,
07
1991
, p. 569–75.
12.
Svensson
,
O.
,
Mathiesen
,
E. B.
,
Reinholt
,
F. P.
, and
Blomgren
,
G.
, “
Formation of a Fulminant Soft-Tissue Pseudotumor After Uncemented Hip Arthroplasty
,”
Journal of Bone and Joint Surgery
, Vol.
70A
, No.
8
,
1988
, pp. 1238–42.
13.
Collier
,
J. P.
,
Surprenant
,
V. A.
,
Jensen
,
R. E.
, and
Mayor
,
M. B.
, “
Corrosion at the Interface of Cobalt-Alloy Heads on Titanium-Alloy Stems
,”
Clinical Orthopedics
, No.
271
,
10
1991
, pp. 305–312.
14.
Flicker
,
D. C.
and
Shivanath
,
R.
, “
Fretting Corrosion Studies of Universal Femoral Head Prostheses and Cone Taper Spigols
,”
Biomaterials
, Vol.
11
,
09
1990
, pp. 495–500.
15.
Bauer
,
T. W.
,
Brown
,
S. A.
,
Jiang
,
M.
, et al
, “
Corrosion in Modular Hip Stems
,”
Transactions of the Orthopaedic Research Society
, Vol.
17
,
1992
, p. 354.
This content is only available via PDF.
You do not currently have access to this chapter.
Close Modal

or Create an Account

Close Modal
Close Modal