Earlier efforts in areas of hip and knee arthroplasties suggest that wear debris, especially from polymeric components, could initiate inflammatory responses leading to peri-prosthetic osteolysis and bone resorption at the implant-bone interface. Aseptic loosening of implants due to particle induced osteolysis is the primary cause of revision surgeries. Metal based implants are considered superior in terms of wear resistance.1–2 However metal-on-metal articulation leads to much smaller sized particulates in comparison to metal-on-polymers. Thus for an equal volume of wear debris from both polymer and metal, the number of metallic particulates can be up to 100 times greater3. Accumulation of metallic debris in the periprosthetic tissue leads to the formation of a fibrous membrane, which might act as a channel for polymeric particulates4. Bench top wear tests as well as bioreactivity studies have emerged as a powerful preclinical tool. However there is still a gap between the in vitro bench-top wear tests and the retrieval test cases. Additionally, these experiments are time consuming, expensive, and labor-intensive procedures. In spite of the fact that experimental data are indispensable, alternatives need to be explored. Predictive finite element modeling based on wear-laws serve as an excellent design tool for parametric analyses. In such models, the effect of individual variables can be judged independently leading to an understanding of the role of that parameter on the final outcome.
Skip Nav Destination
ASME 2011 Summer Bioengineering Conference
June 22–25, 2011
Farmington, Pennsylvania, USA
Conference Sponsors:
- Bioengineering Division
ISBN:
978-0-7918-5458-7
PROCEEDINGS PAPER
The Effect of Different Material Combinations on Wear of an Artificial Cervical Disc as Standalone vs. Placed in a Ligamentous Motion Segment
S. Bhattacharya,
S. Bhattacharya
University of Toledo, Toledo, OH
Search for other works by this author on:
X. Liu
X. Liu
University of Toledo, Toledo, OH
Search for other works by this author on:
S. Bhattacharya
University of Toledo, Toledo, OH
V. K. Goel
University of Toledo, Toledo, OH
A. Kiapour
University of Toledo, Toledo, OH
X. Liu
University of Toledo, Toledo, OH
Paper No:
SBC2011-53027, pp. 559-560; 2 pages
Published Online:
July 17, 2013
Citation
Bhattacharya, S, Goel, VK, Kiapour, A, & Liu, X. "The Effect of Different Material Combinations on Wear of an Artificial Cervical Disc as Standalone vs. Placed in a Ligamentous Motion Segment." Proceedings of the ASME 2011 Summer Bioengineering Conference. ASME 2011 Summer Bioengineering Conference, Parts A and B. Farmington, Pennsylvania, USA. June 22–25, 2011. pp. 559-560. ASME. https://doi.org/10.1115/SBC2011-53027
Download citation file:
7
Views
Related Proceedings Papers
Related Articles
Real-Time, In Vivo Measurement of Contact Pressures at a Knee Arthroplasty
J. Med. Devices (June,2009)
Temporomandibular Joint Disorder and Biomechanical Simulation of the Replacement: A Literature Review on Various Surgical and Nonsurgical Methods, and Development of the Finite Element Method Approach in the Treatment
ASME J of Medical Diagnostics (May,2022)
Design of an Endoreactor for the Cultivation of a Joint-Like-Structure
J. Med. Devices (June,2009)
Related Chapters
Wear and Contact Fatigue Properties of a Novel Lubricant Additive
Bearing and Transmission Steels Technology
Understanding the Problem
Design and Application of the Worm Gear
Cultural Impediments to Practical Modeling of Wear Rates
Tribological Modeling for Mechanical Designers