A relatively recent and non invasive method for characterizing thermal fields generated by high intensity focused ultrasound (HIFU) transducers is Magnetic Resonance (MR) Thermometry method. However, noise signals generated by external RF sources infiltrate the scanner orifice and limit its ability to measure temperature rise during the heating or ablation phase. In this study, MRI monitored HIFU ablations are performed on freshly excised porcine liver samples, at varying sonication times, 20, 30 and 40 s at a constant acoustic intensity level of 1244 W/cm2. Temperature rise during the procedure is measured using Proton Resonant Frequency MR thermometry. Preliminary experiments without an adequate noise filter, failed to record temperature rise during the heating phase. A low pass R-C filter circuit is subsequently incorporated into the experimental set up to prevent infiltration of noise signals in the MRI orifice. This modified RC filter enables measurement of temperature rise during the heating phase followed by temperature decay during cooling. The measured data is within 12% agreement with the temperature rise computed by solving the acoustic and heat equations.
Skip Nav Destination
ASME 2011 Summer Bioengineering Conference
June 22–25, 2011
Farmington, Pennsylvania, USA
Conference Sponsors:
- Bioengineering Division
ISBN:
978-0-7918-5458-7
PROCEEDINGS PAPER
Delineation of Noise Signals From MRI Measured Temperature Rise During HIFU Ablation Procedure
Subhashish Dasgupta,
Subhashish Dasgupta
University of Cincinnati, Cincinnati, OH
Search for other works by this author on:
Seyed Ahmed Dibaji,
Seyed Ahmed Dibaji
University of Cincinnati, Cincinnati, OH
Search for other works by this author on:
Janaka Wansapura,
Janaka Wansapura
Cincinnati Children’s Hospital Medical Center, Cincinnati, OH
Search for other works by this author on:
Matthew R. Myers,
Matthew R. Myers
U. S. Food and Drug Administration, Silver Spring, MD
Search for other works by this author on:
Rupak K. Banerjee
Rupak K. Banerjee
University of Cincinnati, Cincinnati, OH
Search for other works by this author on:
Subhashish Dasgupta
University of Cincinnati, Cincinnati, OH
Seyed Ahmed Dibaji
University of Cincinnati, Cincinnati, OH
Janaka Wansapura
Cincinnati Children’s Hospital Medical Center, Cincinnati, OH
Matthew R. Myers
U. S. Food and Drug Administration, Silver Spring, MD
Rupak K. Banerjee
University of Cincinnati, Cincinnati, OH
Paper No:
SBC2011-53814, pp. 113-114; 2 pages
Published Online:
July 17, 2013
Citation
Dasgupta, S, Dibaji, SA, Wansapura, J, Myers, MR, & Banerjee, RK. "Delineation of Noise Signals From MRI Measured Temperature Rise During HIFU Ablation Procedure." Proceedings of the ASME 2011 Summer Bioengineering Conference. ASME 2011 Summer Bioengineering Conference, Parts A and B. Farmington, Pennsylvania, USA. June 22–25, 2011. pp. 113-114. ASME. https://doi.org/10.1115/SBC2011-53814
Download citation file:
5
Views
Related Proceedings Papers
Related Articles
Reduction of Noise From MR Thermometry Measurements During HIFU Characterization Procedures
J. Nanotechnol. Eng. Med (May,2011)
A New Actuation System With Simulated Electrocardiogram Signal for MR Elastography
J. Med. Devices (June,2010)
Wireless Recharging of Implanted Batteries via Ultrasound
J. Med. Devices (June,2009)
Related Chapters
Introduction and Scope
High Frequency Piezo-Composite Micromachined Ultrasound Transducer Array Technology for Biomedical Imaging
Acoustic Noise in MRI Scanners
Biomedical Applications of Vibration and Acoustics in Therapy, Bioeffect and Modeling
EVALUATION OF THERMAL PROBE COOLING EFFECT BASED ON PIPELINE PARALLEL OPTICAL CABLE
Pipeline Integrity Management Under Geohazard Conditions (PIMG)