In this work, we present a 2D numerical flow study to analyze the impact of different stent strut profiles on the shear stress distribution for stented coronary arteries. In that context, five cross-sectional profiles were considered: square, circular, elliptical, and a two tear-drop shapes. Specific performance metrics based on the statistical moments were developed to assess the variation of shear stress distributions along the wall with respected to a normal reference condition (non-stented segment). The results show that stent strut profile significantly affects the shear stress distribution along the wall. It also demonstrate that more streamline profiles such as the tear-drop and elliptical profiles exhibit better performance than the standard square and circular profiles for shear stress.
Skip Nav Destination
ASME 2007 Summer Bioengineering Conference
June 20–24, 2007
Keystone, Colorado, USA
Conference Sponsors:
- Bioengineering Division
ISBN:
0-7918-4798-5
PROCEEDINGS PAPER
Evaluation of the Effect of Stent Strut Profile on Shear Stress Distribution
Bilal Ruzzeh,
Bilal Ruzzeh
McGill University, Montreal, QC, Canada
Search for other works by this author on:
Rosaire Mongrain,
Rosaire Mongrain
McGill University, Montreal, QC, Canada
Search for other works by this author on:
Richard Leask,
Richard Leask
McGill University, Montreal, QC, Canada
Search for other works by this author on:
Olivier Bertrand
Olivier Bertrand
Laval Hospital
Search for other works by this author on:
Bilal Ruzzeh
McGill University, Montreal, QC, Canada
Rosaire Mongrain
McGill University, Montreal, QC, Canada
Richard Leask
McGill University, Montreal, QC, Canada
Olivier Bertrand
Laval Hospital
Paper No:
SBC2007-177056, pp. 773-774; 2 pages
Published Online:
March 12, 2014
Citation
Ruzzeh, B, Mongrain, R, Leask, R, & Bertrand, O. "Evaluation of the Effect of Stent Strut Profile on Shear Stress Distribution." Proceedings of the ASME 2007 Summer Bioengineering Conference. ASME 2007 Summer Bioengineering Conference. Keystone, Colorado, USA. June 20–24, 2007. pp. 773-774. ASME. https://doi.org/10.1115/SBC2007-177056
Download citation file:
8
Views
Related Proceedings Papers
Related Articles
Two-Dimensional Optimization of a Stent for an Aneurysm
J. Med. Devices (June,2010)
Coronary Artery Stenting Affects Wall Shear Stress Topological Skeleton
J Biomech Eng (June,2022)
The Localized Hemodynamics of Drug-Eluting Stents Are Not Improved by the Presence of Magnetic Struts
J Biomech Eng (January,2017)
Related Chapters
mDFA Human Empirical Results
Modified Detrended Fluctuation Analysis (mDFA)
In-Vitro Modeling of the Dynamic Forces in the Femoropopliteal Artery
Fatigue and Fracture of Medical Metallic Materials and Devices: 2nd Volume
The Quest for Fatigue-Resistant Nitinol for Medical Implants
Fourth Symposium on Fatigue and Fracture of Metallic Medical Materials and Devices