In this paper, performance of aortic heart valve prosthesis in different geometries of the aortic root is investigated experimentally. The objective of this investigation is to establish a set of parameters, which are associated with abnormal flow patterns due to the flow through a prosthetic heart valve implanted to the patients that had certain types of valve diseases prior to the valve replacement. Specific valve diseases, classified into two clinical categories, were correlated with the corresponding changes of aortic root geometry. These categories correspond to aortic valve stenosis and aortic valve insufficiency. The control case that corresponds to the aortic root of a patient without valve disease was used as a reference. Experiments were performed at test conditions corresponding to 70 beats/min, 5.5 L/min target cardiac output and a mean aortic pressure of 100 mmHg. By varying the aortic root geometry, it was possible to investigate corresponding changes in the levels of Reynolds shear stress and establish the possibility of platelet activation and, as a result of that, the formation of blood clots.
Skip Nav Destination
ASME 2013 Pressure Vessels and Piping Conference
July 14–18, 2013
Paris, France
Conference Sponsors:
- Pressure Vessels and Piping Division
- Nondestructive Evaluation Engineering Division
ISBN:
978-0-7918-5568-3
PROCEEDINGS PAPER
In Vitro Study of the Influence of the Aortic Root Geometry on Flow Characteristics of a Prosthetic Heart Valve
Oleksandr Barannyk,
Oleksandr Barannyk
University of Victoria, Victoria, BC, Canada
Search for other works by this author on:
Satya Karri,
Satya Karri
ViVitro Labs, Inc., Victoria, BC, Canada
Search for other works by this author on:
Peter Oshkai
Peter Oshkai
University of Victoria, Victoria, BC, Canada
Search for other works by this author on:
Oleksandr Barannyk
University of Victoria, Victoria, BC, Canada
Satya Karri
ViVitro Labs, Inc., Victoria, BC, Canada
Peter Oshkai
University of Victoria, Victoria, BC, Canada
Paper No:
PVP2013-97105, V004T04A002; 10 pages
Published Online:
January 17, 2014
Citation
Barannyk, O, Karri, S, & Oshkai, P. "In Vitro Study of the Influence of the Aortic Root Geometry on Flow Characteristics of a Prosthetic Heart Valve." Proceedings of the ASME 2013 Pressure Vessels and Piping Conference. Volume 4: Fluid-Structure Interaction. Paris, France. July 14–18, 2013. V004T04A002. ASME. https://doi.org/10.1115/PVP2013-97105
Download citation file:
11
Views
Related Articles
In Vitro Thrombogenicity Testing of Biomaterials in a Dynamic Flow Loop: Effects of Length and Quantity of Test Samples
J. Med. Devices (September,2023)
Two-Dimensional Simulation of Flow and Platelet Dynamics in the Hinge Region of a Mechanical Heart Valve
J Biomech Eng (March,2009)
Flow and Thrombosis at Orifices Simulating Mechanical Heart Valve Leakage Regions
J Biomech Eng (February,2006)
Related Chapters
Introduction
Mechanical Blood Trauma in Circulatory-Assist Devices
Dynamic Behavior of Pumping Systems
Pipeline Pumping and Compression Systems: A Practical Approach
A Simple Carburetor
Case Studies in Fluid Mechanics with Sensitivities to Governing Variables