Abstract

“Zero-drift” characteristics of an optical intraocular pressure sensor is investigated as a function of membrane dissolution and hygroscopic swelling of the epoxy layer. Both effects were studied using an analytical model based on the deflection of a circular membrane. Results from the analytical model were verified with experimental results from “bulge” testing. The analytical model was used to study the “zero drift” of the sensor as a function of changes in membrane thickness and geometry of the sensor. The results show that dissolution of the membrane and swelling of the spacer layer can contribute to zero-drift over time. The results are useful in guiding design and fabrication optimization to minimize drift in intraocular pressure sensors used for long term implantation.

This content is only available via PDF.
You do not currently have access to this content.