Radiofrequency (RF) catheter ablation is a highly effective treatment for many cardiac arrhythmias, especially for tachyarrhythmia. RF energy is introduced through the catheter onto the endocardial surface to destroy the abnormal heart tissue causing the heart rhythm disorder. Many parameters relate to myocardial temperature, such as RF power, tissue contact, and blood flow. Blood flow is an important factor that has a cooling effect on myocardium and affects the final lesion size. Many previous studies have shown that under temperature control, lesion sizes are larger and tissue temperatures rise faster with a high flow rate. If the flow causes a decrease in the temperature of the catheter tip, the generator will increase the power output to maintain the tip at a constant temperature. However, few studies of RF catheter ablation have investigated how ablation affects blood flow. Observation of the flow pattern around the catheter can help to determine the mechanism of the flow effects on the temperature of the catheter tip. The purpose of this study is to observe the flow pattern during ablation using an in-vitro circulation system developed for Particle Image Velocimetry (PIV). We developed an open-channel circulation system to simulate blood flow. The mold for the open-channel was built with acrylic boards for transparency. The working fluid was 0.9% saline, which was used at room temperature (20°C). Instead of animal myocardium, we used a poly (vinyl alcohol) hydrogel (PVA-H), which has mechanical characteristics that approximate those of biological soft tissue, and contact with the PVA-H surface by the catheter is similar to that with myocardium. A 7 Fr catheter with a 4-mm ablation electrode tip was set perpendicular to the PVA-H surface, and the contact weight between the electrode of the catheter and the PVA-H surface was 2.2 gf. To measure the temperature profile in the PVA-H, a K-type thermocouple with the diameter of 0.5 mm was placed at the depth of 2 mm from the surface. The thermocouple tip was always placed on the catheter axis. The flow pattern at the location where the catheter was held was observed by a high speed camera, and the resulting images were analyzed by particle image velocimetry (PIV). The results showed that in the absence of applied flow, convection flow from the electrode is observed in the areas around the catheter. However, under a 1.6 L/min flow rate, convection flow disappears. In conclusion, blood flow could decrease the catheter tip temperature, and the influence of ablation in the flow around the catheter disappeared.
Skip Nav Destination
ASME 2013 International Mechanical Engineering Congress and Exposition
November 15–21, 2013
San Diego, California, USA
Conference Sponsors:
- ASME
ISBN:
978-0-7918-5621-5
PROCEEDINGS PAPER
PIV Analysis of the Flow Pattern Around an Ablation Catheter to Observe the Flow Effect on the Electrode
Kaihong Yu,
Kaihong Yu
Tohoku University, Sendai, Miyagi, Japan
Search for other works by this author on:
Tetsui Yamashita,
Tetsui Yamashita
JMS Co., Ltd., Tokyo, Japan
Search for other works by this author on:
Shigeaki Shingyochi,
Shigeaki Shingyochi
Nidec Copal Electronics Corp., Sano, Tochigi, Japan
Search for other works by this author on:
Kazuo Matsumoto,
Kazuo Matsumoto
Saitama Medical University, Hidaka, Saitama, Japan
Search for other works by this author on:
Makoto Ohta
Makoto Ohta
Tohoku University, Sendai, Miyagi, Japan
Search for other works by this author on:
Kaihong Yu
Tohoku University, Sendai, Miyagi, Japan
Tetsui Yamashita
JMS Co., Ltd., Tokyo, Japan
Shigeaki Shingyochi
Nidec Copal Electronics Corp., Sano, Tochigi, Japan
Kazuo Matsumoto
Saitama Medical University, Hidaka, Saitama, Japan
Makoto Ohta
Tohoku University, Sendai, Miyagi, Japan
Paper No:
IMECE2013-62791, V03AT03A025; 9 pages
Published Online:
April 2, 2014
Citation
Yu, K, Yamashita, T, Shingyochi, S, Matsumoto, K, & Ohta, M. "PIV Analysis of the Flow Pattern Around an Ablation Catheter to Observe the Flow Effect on the Electrode." Proceedings of the ASME 2013 International Mechanical Engineering Congress and Exposition. Volume 3A: Biomedical and Biotechnology Engineering. San Diego, California, USA. November 15–21, 2013. V03AT03A025. ASME. https://doi.org/10.1115/IMECE2013-62791
Download citation file:
12
Views
Related Proceedings Papers
Related Articles
CT Visualization of Cryoablation in Pulmonary Veins
J. Med. Devices (June,2009)
Benchmark Solution for the Prediction of Temperature Distributions During Radiofrequency Ablation of Cardiac Tissue
J Biomech Eng (August,2004)
Design of a Cooling Guide Catheter for Rapid Heart Cooling
J. Med. Devices (September,2010)
Related Chapters
Introduction and Scope
High Frequency Piezo-Composite Micromachined Ultrasound Transducer Array Technology for Biomedical Imaging
Completing the Picture
Air Engines: The History, Science, and Reality of the Perfect Engine
Even a Watched Pot Boils Eventually
Hot Air Rises and Heat Sinks: Everything You Know about Cooling Electronics Is Wrong