The reasons leading to the in vivo failure of three titanium alloy modular implants in cementless total hip arthroplasty were investigated. The implantation period ranged from 18 months to over three years. Two were fractured in vivo and the other was retrieved secondary to aseptic loosening after 18 months in service. Macroscopic examination showed close topographical similarities between the two fractured implants. Dark elliptical striations on the fractured site indicated the occurrence of low cycle fatigue. Light Microscopy and Scanning Electron Microscopy confirmed the presence of fretting, pitting, plastic deformation, and stress-induced corrosion cracking. In two of the three implants, EDS confirmed metal ion traces in the tissue around the implant. However, nothing unusual was found in the third unfractured implant. Taper performance is influenced by metallurgy, the load carried and the effect of the local microenvironment. Methods to reduce the impact of these factors may reduce the incidence of taper related failure.
Skip Nav Destination
ASME 2008 International Mechanical Engineering Congress and Exposition
October 31–November 6, 2008
Boston, Massachusetts, USA
Conference Sponsors:
- ASME
ISBN:
978-0-7918-4863-0
PROCEEDINGS PAPER
Failure of Three Cementless Modular Total Hip Arthroplasty Prostheses: A Retrieval Analysis
Manish Paliwal,
Manish Paliwal
The College of New Jersey, Ewing, NJ
Search for other works by this author on:
D. Gordon Allan,
D. Gordon Allan
Southern Illinois University School of Medicine, Springfield, IL
Search for other works by this author on:
Peter Filip
Peter Filip
Southern Illinois University - Carbondale, Carbondale, IL
Search for other works by this author on:
Manish Paliwal
The College of New Jersey, Ewing, NJ
D. Gordon Allan
Southern Illinois University School of Medicine, Springfield, IL
Peter Filip
Southern Illinois University - Carbondale, Carbondale, IL
Paper No:
IMECE2008-66998, pp. 97-105; 9 pages
Published Online:
August 26, 2009
Citation
Paliwal, M, Allan, DG, & Filip, P. "Failure of Three Cementless Modular Total Hip Arthroplasty Prostheses: A Retrieval Analysis." Proceedings of the ASME 2008 International Mechanical Engineering Congress and Exposition. Volume 2: Biomedical and Biotechnology Engineering. Boston, Massachusetts, USA. October 31–November 6, 2008. pp. 97-105. ASME. https://doi.org/10.1115/IMECE2008-66998
Download citation file:
12
Views
Related Proceedings Papers
Related Articles
Microstructural Development and Cracking Behavior of AISI 304 Stainless Steel Tested in Time Dependent Fatigue Modes
J. Eng. Mater. Technol (January,1983)
Effect of dc on the Formability of Ti–6Al–4V
J. Eng. Mater. Technol (July,2009)
Design of an Endoreactor for the Cultivation of a Joint-Like-Structure
J. Med. Devices (June,2009)
Related Chapters
Surface Analysis and Tools
Tribology of Mechanical Systems: A Guide to Present and Future Technologies
Understanding the Problem
Design and Application of the Worm Gear
Introduction
A Practical Guide to Avoiding Steam Purity Problems in the Industrial Plant (CRTD-35)