To improve treatment planning in boron neutron capture therapy (BNCT), a new Monte-Carlo radiotherapy planning system is under development at Japan Atomic Energy Agency (JAEA). This system (developing code: JCDS-FX) builds on fundamental technologies of JCDS (JAEA Computation Dosimetry System) which has been applied to actual BNCT trials at Japan Research Reactor No.4 (JRR-4). Basic technologies of JCDS have been taken over to JCDS-FX, and some new functions have been built into the system. One of features of the JCDS-FX is that PHITS as a multi-purpose particle Monte-Carlo transport code has been applied to particle transport calculation. Application of PHITS enables to evaluate doses for neutrons and photons as well as protons and heavy ions. Therefore, the JCDS-FX with PHITS can perform treatment planning for not only BNCT but also particle radiotherapy. To verify calculation accuracy of the JCDS-FX with PHITS, dose evaluations for neutron irradiation of a cylindrical water phantom and for an actual clinical trial conducted at JRR-4 were performed. The verification results indicated that JCDS-FX is applicable to BNCT treatment planning in practical use. Further verifications of the system are being performed to achieve practical application of the system in the future. And in addition to the BNCT, investigations for application of the system to any other particle radiotherapy like proton therapy are carried forward.
Skip Nav Destination
17th International Conference on Nuclear Engineering
July 12–16, 2009
Brussels, Belgium
Conference Sponsors:
- Nuclear Engineering Division
ISBN:
978-0-7918-4352-9
PROCEEDINGS PAPER
Development of a New Monte-Carlo Treatment Planning System for Boron Neutron Capture Therapy
Hiroaki Kumada,
Hiroaki Kumada
Japan Atomic Energy Agency, Tokai, Ibaraki, Japan
Search for other works by this author on:
Takemi Nakamura,
Takemi Nakamura
Japan Atomic Energy Agency, Tokai, Ibaraki, Japan
Search for other works by this author on:
Akira Matsumura,
Akira Matsumura
University of Tsukuba, Tsukuba, Ibaraki, Japan
Search for other works by this author on:
Koji Ono
Koji Ono
Kyoto University, Kumatori, Osaka, Japan
Search for other works by this author on:
Hiroaki Kumada
Japan Atomic Energy Agency, Tokai, Ibaraki, Japan
Takemi Nakamura
Japan Atomic Energy Agency, Tokai, Ibaraki, Japan
Akira Matsumura
University of Tsukuba, Tsukuba, Ibaraki, Japan
Koji Ono
Kyoto University, Kumatori, Osaka, Japan
Paper No:
ICONE17-75407, pp. 823-829; 7 pages
Published Online:
February 25, 2010
Citation
Kumada, H, Nakamura, T, Matsumura, A, & Ono, K. "Development of a New Monte-Carlo Treatment Planning System for Boron Neutron Capture Therapy." Proceedings of the 17th International Conference on Nuclear Engineering. Volume 2: Structural Integrity; Safety and Security; Advanced Applications of Nuclear Technology; Balance of Plant for Nuclear Applications. Brussels, Belgium. July 12–16, 2009. pp. 823-829. ASME. https://doi.org/10.1115/ICONE17-75407
Download citation file:
5
Views
Related Proceedings Papers
Related Articles
Vision-Based Patient Body Tracking in Helical Tomotherapy
J. Med. Devices (June,2008)
A Novel Translational Total Body Irradiation Technique
J. Med. Devices (September,2010)
The Use of Polycarbonate as Dosimeter of High Doses
ASME J of Nuclear Rad Sci (April,2021)
Related Chapters
Validation of Elekta iViewGT A-Si EPID Model for Pre-Treatment Dose Verification of IMRT Fields
International Conference on Advanced Computer Theory and Engineering, 4th (ICACTE 2011)
New Generation Reactors
Energy and Power Generation Handbook: Established and Emerging Technologies
First Results from a Probabilistic Risk Assessment for PSI's Spot-Scanning Proton Therapy Facility (PSAM-0390)
Proceedings of the Eighth International Conference on Probabilistic Safety Assessment & Management (PSAM)