Thermal ablation is rapidly becoming a standard of care for the treatment of atrial fibrillation (AF), a cardiac disorder characterized by irregular heart rhythm and estimated to impact more than 33 million people worldwide [1]. AtriCure is a company that specializes in epicardial ablation for AF and here we describe the development of a numerical model to study the performance of the Isolator® Synergy™ Clamp bipolar radiofrequency (RF) device. The clamp device features two jaws with embedded electrode pairs, which are used to secure the tissue by clamping across the left atrium (as shown in Figure 1). Energy is applied between the bipolar electrodes at approximately 460 kHz through an impedance-based control algorithm and is additionally duty-cycled between the pairs to further distribute the heating. Patient anatomies vary greatly and measured impedance will depend on atrial wall thickness, epicardial fat, electrode-tissue engagement, and structural variations. Further, tissue conductivity (inversely related to impedance) increases as the tissue is heated, leading to a complicated process, where the heat generation depends on the impedance, which in turn is a strong function of temperature. Energy delivery continues until a phase change in the tissue’s water content occurs, producing a sharp increase in impedance and termination of the ablation. Therefore, since tissue impedance and heating drive the device’s performance, a majority of the effort described here focuses on the validation work done to ensure the model is based on an accurate description of the tissue properties and response. While previous modeling of RF ablation often does include temperature-dependence of tissue properties, the referenced values vary notably and rarely include direct validation of modeling results to benchtop data. Variations in anatomy and fat content can dramatically impact the energy delivery and patient-to-patient treatment efficacy, so an accurate description of the tissue response is critical to understanding the limitations of current energy delivery algorithms and provides an invaluable tool in designing more efficacious ablation devices and algorithms.
Skip Nav Destination
2018 Design of Medical Devices Conference
April 9–12, 2018
Minneapolis, Minnesota, USA
ISBN:
978-0-7918-4078-8
PROCEEDINGS PAPER
Development and Validation of Numerical Model Simulation for RF Ablation Using the Isolator Synergy Clamp
Michael Etheridge,
Michael Etheridge
AtriCure Inc., Minnetonka, MN
Search for other works by this author on:
Harishankar Natesan,
Harishankar Natesan
AtriCure Inc., Minnetonka, MN
Search for other works by this author on:
Radek Lopusnik,
Radek Lopusnik
AtriCure Inc., Minnetonka, MN
Search for other works by this author on:
Adam Cates
Adam Cates
AtriCure Inc., Minnetonka, MN
Search for other works by this author on:
Michael Etheridge
AtriCure Inc., Minnetonka, MN
Harishankar Natesan
AtriCure Inc., Minnetonka, MN
Radek Lopusnik
AtriCure Inc., Minnetonka, MN
Adam Cates
AtriCure Inc., Minnetonka, MN
Paper No:
DMD2018-6949, V001T01A017; 4 pages
Published Online:
June 14, 2018
Citation
Etheridge, M, Natesan, H, Lopusnik, R, & Cates, A. "Development and Validation of Numerical Model Simulation for RF Ablation Using the Isolator Synergy Clamp." Proceedings of the 2018 Design of Medical Devices Conference. 2018 Design of Medical Devices Conference. Minneapolis, Minnesota, USA. April 9–12, 2018. V001T01A017. ASME. https://doi.org/10.1115/DMD2018-6949
Download citation file:
300
Views
Related Proceedings Papers
Related Articles
Benchmark Solution for the Prediction of Temperature Distributions During Radiofrequency Ablation of Cardiac Tissue
J Biomech Eng (August,2004)
A Comparative Modeling Study of Thermal Mitigation Strategies in Irreversible Electroporation Treatments
J. Heat Transfer (March,2022)
Finite Element Simulation of Proportional, Integral, and Derivative-Controlled Bipolar Radiofrequency Ablation of Porcine Spinal Muscle
ASME J of Medical Diagnostics (May,2023)
Related Chapters
Experimental Investigation of an Improved Thermal Response Test Equipment for Ground Source Heat Pump Systems
Inaugural US-EU-China Thermophysics Conference-Renewable Energy 2009 (UECTC 2009 Proceedings)
Discussion of Parameters to Be Monitored
Consensus on Operating Practices for the Sampling and Monitoring of Feedwater and Boiler Water Chemistry in Modern Industrial Boilers (CRTD-81)
Threshold Functions
Closed-Cycle Gas Turbines: Operating Experience and Future Potential