Thrombogenicity testing continues to be a critical requirement for regulatory approval of blood-contacting medical devices and the ISO guidelines have recently been updates [1]. This new guideline ascribes value to both in vivo and in vitro testing including both the non-anticoagulated venous implant (NAVI) model, and the new methods for in vitro testing. One challenge with the animal-blood-based in vitro assays that have been validated and used for submissions is that they still may not accurately translate to clinical safety or predict the risk for thrombogenic potential in humans. We have previously described a model using minimally heparinized ovine blood and are continuing to improve the overall methodology [2,3]. In addition, we have transferred these methods to a human blood assay which therefore has enhanced potential for prediction of clinical risk. As with the ovine model, the key characteristics of a successful in vitro method include fresh blood, low levels of anticoagulation, flow conditions and minimization of air/blood interfaces. This human model integrates freshly harvested human blood containing minimal levels of heparin with variable flow from a unidirectional peristaltic pump and unlike many of the human blood assays, it can accommodate larger devices and higher flow rates than previously described [1,4]. Control materials which were optimized in the ovine model were also used to reproducibly elicit positive and negative thrombogenic responses. We feel that this model can be used for validation of the ovine model with cross comparisons of a number of legally marketed comparator devices. Alternatively, if the human blood methodology can be streamlined and performed cost effectively on a regular and basis, this assay could supplant the current ovine model and allow a highly predictive preclinical test for thrombogenicity of devices.
Skip Nav Destination
2018 Design of Medical Devices Conference
April 9–12, 2018
Minneapolis, Minnesota, USA
ISBN:
978-0-7918-4078-8
PROCEEDINGS PAPER
Thrombogenicity Testing for Blood-Contacting Medical Devices in an In Vitro Human Blood-Loop
Matt Cunningham,
Matt Cunningham
American Preclinical Services, Minneapolis, MN
Search for other works by this author on:
Sarah Howard,
Sarah Howard
American Preclinical Services, Minneapolis, MN
Search for other works by this author on:
Abby Beltrame,
Abby Beltrame
American Preclinical Services, Minneapolis, MN
Search for other works by this author on:
Yan Chen,
Yan Chen
American Preclinical Services, Minneapolis, MN
Search for other works by this author on:
Mark Smith
Mark Smith
American Preclinical Services, Minneapolis, MN
Search for other works by this author on:
Matt Cunningham
American Preclinical Services, Minneapolis, MN
Sarah Howard
American Preclinical Services, Minneapolis, MN
Abby Beltrame
American Preclinical Services, Minneapolis, MN
Yan Chen
American Preclinical Services, Minneapolis, MN
Mark Smith
American Preclinical Services, Minneapolis, MN
Paper No:
DMD2018-6875, V001T01A007; 3 pages
Published Online:
June 14, 2018
Citation
Cunningham, M, Howard, S, Beltrame, A, Chen, Y, & Smith, M. "Thrombogenicity Testing for Blood-Contacting Medical Devices in an In Vitro Human Blood-Loop." Proceedings of the 2018 Design of Medical Devices Conference. 2018 Design of Medical Devices Conference. Minneapolis, Minnesota, USA. April 9–12, 2018. V001T01A007. ASME. https://doi.org/10.1115/DMD2018-6875
Download citation file:
764
Views
Related Proceedings Papers
Related Articles
Thrombogenicity Testing of Medical Devices in a Minimally Heparinized Ovine Blood Loop
J. Med. Devices (June,2017)
Harnessing Experience for Efficient Medical Device Product Development
J. Med. Devices (June,2009)
Ventricular Assist Devices: Current State and Challenges
J. Med. Devices (December,2017)
Related Chapters
QRAS Approach to Phased Mission Analysis (PSAM-0444)
Proceedings of the Eighth International Conference on Probabilistic Safety Assessment & Management (PSAM)
Specific Device Examples
Magnetic Bearings for Mechanical Cardiac Assist Devices
Concluding remarks
Mechanical Blood Trauma in Circulatory-Assist Devices